The Allied High Tech MetPrep 4 is a microprocessor-controlled grinding/polishing machine for semiautomatic sample preparation. The MetPrep 4 is programmable for up to 25 steps, including platen speed, comp/contra rotation, cycle time, fluid selection, sample force, sample rotation speed, platen rinse and force reduction. The MetPrep 4 is designed and manufactured by Allied in the USA and carries a two-year warranty.[1][2]

Allied High TechMetprep 4
No photo available yet

What it is

The MetPrep 4 is a semiautomatic grinding and polishing machine manufactured by Allied High Tech Products. The system can be combined with PH-4 or PH-6 power heads for low-to-high volume sample preparation. The MetPrep 4 is designed and manufactured in the United States.[1][4]

How it works

The MetPrep 4 uses a rotating platen and a pneumatically or electrically actuated power head that holds sample mounts. The power head applies controlled force to the samples against the abrasive surface. The microprocessor-based controller allows programming of up to 25 steps, including parameters for platen speed, rotation direction, cycle time, fluid selection, sample force, sample rotation speed, water rinse, and force reduction.[1]

The system supports either central force mode, where multiple samples are pressed together on a common plane, or individual force mode, where each sample receives independent force. The power head tilts up to allow unhindered access to the platen and sample holder. The platen can be aluminum or magnetic, and the machine includes an electronic coolant control with adjustable flow, a bowl flush to remove debris, and an emergency shut-off switch.[1]

The MetPrep 4 uses a three-phase AC motor with single-phase input and a reduction gearbox for constant high-torque output. For 115 volt models, platen speed ranges from 50 to 400 RPM in 10 RPM increments; for 230 volt models, the range is 40 to 600 RPM. Jog speed has the same ranges. The platen diameters available are 10 inches (250 mm) or 12 inches (300 mm).[1]

What do the numbers mean?

Power & electrical16

Platen speed, 115 V
50-400 RPM[1]
Accurate?
Platen speed, 230 V
40-600 RPM[1]
Accurate?
Platen jog speed, 115 V
50-400 RPM[1]
Accurate?
Platen jog speed, 230 V
40-600 RPM[1]
Accurate?
Horsepower, 115 V
1.5 HP (1100 W)[1]
Accurate?
Horsepower, 230 V
2 HP (1500 W)[1]
Accurate?
Motor
3-phase AC motor, single-phase AC input, with durable reduction gearbox providing constant high-torque output[1]
Accurate?
Platen Speed (115 V)
50-400 RPM (10 RPM increments)[1]
Accurate?
Platen Speed (230 V)
40-600 RPM (10 RPM increments)[1]
Accurate?
Jog Speed (115 V)
50-400 RPM (10 RPM increments)[1]
Accurate?
Jog Speed (230 V)
40-600 RPM (10 RPM increments)[1]
Accurate?
Motor Type
3 Phase AC motor, single phase AC input with reduction gearbox[1]
Accurate?
Horsepower (115 V)
1.5 HP[1]
Accurate?
Horsepower (230 V)
2 HP[1]
Accurate?
Power (115 V)
1100 W[1]
Accurate?
Power (230 V)
1500 W[1]
Accurate?

Control & software2

Interface
Touch-pad switches and backlit 4-line LCD[1]
Accurate?
Coolant
Electronic coolant control with adjustable flow control valve; adjustable pivoting coolant nozzle allows sample/bowl cleaning[4]
Accurate?

Dimensions & facilities3

Dimensions
22" (560 mm) W x 27" (690 mm) D x 24" (610 mm) H[1]
Accurate?
Weight
225 lb. (102 kg)[1]
Accurate?
Dimensions (W x D x H)
22 x 27 x 24 inches (560 x 690 x 610 mm)[1]
Accurate?

Configuration & options26

With PH 4 Pow3er Head and AD 5 Fluid Dispenser[3]
Accurate?
Platen diameter
10" (250 mm) or 12" (300 mm)[1]
Accurate?
Programmable steps
Up to 25[1]
Accurate?
Cycle time
0-60 minutes in 15-second increments (brochure); 0-120 minutes in 15-second increments (Ellipsiz)[1]
Accurate?
Sample rotation speed (PH-4/PH-6 power head)
0-150 RPM in 10 RPM increments[1]
Accurate?
PH-4 individual force sample count
1-4 samples[1]
Accurate?
PH-6 individual force sample count
1-6 samples[1]
Accurate?
PH-4 central force sample count
3-9 samples[1]
Accurate?
PH-6 central force sample count
3-12 samples[1]
Accurate?
PH-4/PH-6 central force range
5-90 LbF / 22-394 N[1]
Accurate?
PH-4/PH-6 individual force range
0-22 LbF / 0-96 N[1]
Accurate?
Maximum mount size (PH-4/PH-6)
2" / 51 mm[1]
Accurate?
Warranty
Two (2) year warranty[1]
Accurate?
Manufacturer
Designed & manufactured by Allied in the USA[1]
Accurate?
Compliance
CE compliant for EU[4]
Accurate?
Construction
Sturdy RIM, aluminum and stainless steel construction with low-profile design; molded one-piece corrosion-resistant cover[4]
Accurate?
Platen Diameter
10 inch (250 mm) or 12 inch (300 mm)[1]
Accurate?
Cycle Time
0-60 minutes (15 second increments)[1]
Accurate?
Sample Rotation Speed (PH-4/PH-6)
0-150 RPM (10 RPM increments)[1]
Accurate?
Sample Force Range (PH-4/PH-6, Individual Force Mode)
0-22 LbF (0-96 N)[1]
Accurate?
Sample Force Range (PH-4/PH-6, Central Force Mode)
5-90 LbF (22-394 N)[1]
Accurate?
Maximum Mount Size
2 inch (51 mm)[1]
Accurate?
Number of Samples (PH-4 Individual Force)
1-4[1]
Accurate?
Number of Samples (PH-6 Individual Force)
1-6[1]
Accurate?
Number of Samples (PH-4 Central Force)
3-9[1]
Accurate?
Number of Samples (PH-6 Central Force)
3-12[1]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
MetPrep 4 with PH-4 Power Head——Combines the MetPrep 4 with the PH-4 power head; system item numbers 5-6510 (10" platen, 115 V), 5-6510-230 (10" platen, 230 V), 5-6512 (12" platen, 115 V), 5-6512-230 (12" platen, 230 V).alliedhightech.com[1]
MetPrep 4 with PH-6 Power Head——Combines the MetPrep 4 with the PH-6 power head; system item numbers 5-6610 (10" platen, 115 V), 5-6610-230 (10" platen, 230 V), 5-6612 (12" platen, 115 V), 5-6612-230 (12" platen, 230 V). The PH-6 maximizes use of the entire 12" abrasive disc when using individual force mode.alliedhightech.com[1]
5-6510 (MetPrep 4 with PH-4, 10" platen, 115 V)——10-inch platen with PH-4 Power Head, 115 VAC inputalliedhightech.com[1]
5-6610 (MetPrep 4 with PH-6, 10" platen, 115 V)——10-inch platen with PH-6 Power Head, 115 VAC inputalliedhightech.com[1]
5-6512 (MetPrep 4 with PH-4, 12" platen, 115 V)——12-inch platen with PH-4 Power Head, 115 VAC inputalliedhightech.com[1]
5-6612 (MetPrep 4 with PH-6, 12" platen, 115 V)——12-inch platen with PH-6 Power Head, 115 VAC inputalliedhightech.com[1]
5-6510-230 (MetPrep 4 with PH-4, 10" platen, 230 V)——10-inch platen with PH-4 Power Head, 230 VAC inputalliedhightech.com[1]
5-6610-230 (MetPrep 4 with PH-6, 10" platen, 230 V)——10-inch platen with PH-6 Power Head, 230 VAC inputalliedhightech.com[1]
5-6512-230 (MetPrep 4 with PH-4, 12" platen, 230 V)——12-inch platen with PH-4 Power Head, 230 VAC inputalliedhightech.com[1]
5-6612-230 (MetPrep 4 with PH-6, 12" platen, 230 V)——12-inch platen with PH-6 Power Head, 230 VAC inputalliedhightech.com[1]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Platen speed, 115 V50-400 RPM[1]
  • Platen speed, 230 V40-600 RPM[1]
  • Platen jog speed, 115 V50-400 RPM[1]
  • Platen jog speed, 230 V40-600 RPM[1]
  • Horsepower, 115 V1.5 HP (1100 W)[1]
  • Horsepower, 230 V2 HP (1500 W)[1]
  • Motor3-phase AC motor, single-phase AC input, with durable reduction gearbox providing constant high-torque output[1]
  • Platen Speed (115 V)50-400 RPM (10 RPM increments)[1]
  • Platen Speed (230 V)40-600 RPM (10 RPM increments)[1]
  • Jog Speed (115 V)50-400 RPM (10 RPM increments)[1]
  • Jog Speed (230 V)40-600 RPM (10 RPM increments)[1]
  • Motor Type3 Phase AC motor, single phase AC input with reduction gearbox[1]
  • Horsepower (115 V)1.5 HP[1]
  • Horsepower (230 V)2 HP[1]
  • Power (115 V)1100 W[1]
  • Power (230 V)1500 W[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

How many programmable steps does the Metprep 4 support?

The Metprep 4 supports up to twenty-five programmable steps.[1][2]

What power-head configurations are used with the Metprep 4?

The Metprep 4 is used with the PH-4 or PH-6 power head.[1][2]

What force modes does the system support?

The system supports central force and individual force modes.[1][2]

What platen sizes are specified for the Metprep 4?

The Metprep 4 is specified with ten inch and twelve inch platens.[1][2]

What sample workload range is the Metprep 4 intended for?

The Metprep 4 is intended for low to high volume sample preparation.[1][4][2]

Not publicly documented

The following facts about the Metprep 4 are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the Metprep 4 are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the Metprep 4 are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the Metprep 4 are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the Metprep 4 is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the Metprep 4 are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (4)Every fact above is drawn from these public sources
  1. [1]alliedhightech.com — alliedhightech.comalliedhightech.com
  2. [2]MetPrep 4 Fly Sheet - Allied High Tech Products Inc. — doczz.netdoczz.net
  3. [3]inventory listing
  4. [4]ellipsizdss.com — ellipsizdss.comellipsizdss.com
Was this page accurate?

Last updated Oct 1, 2026.

Compare with similar tools

View all →